Comparison of ejection fraction calculation between CT and SPECT at high heart rate: A dynamic cardiac phantom study.
Ontology highlight
ABSTRACT: The purpose of this study is to compare the ejection fraction (EF) calculation of CT and SPECT at high heart rate. A dynamic cardiac phantom with programmable end-systolic volume (ESV), end-diastolic volume (EDV), and heart rate was used to compare CT, which has high spatial resolution (< 1 mm) and modest temporal resolution of 175 msec, and SPECT, which has high temporal resolution of 16 bins per cardiac cycle but poor spatial resolution (> 1 cm) in EF, ESV, and EDV at the heart rates ≤ 100 bpm for EF = 30 (disease state) and EF = 60 (healthy state). EF calculations for SPECT were accurate in 2% for 40 to 100 bpm for both EF = 30 and EF = 60, and were not heart rate dependent although both ESV and EDV could be underestimated by 18-20%. EF calculations for CT were accurate in 2.2% for 40 a
SUBMITTER: Chen CC
PROVIDER: S-EPMC10905667 | biostudies-literature | 2021 Feb