ABSTRACT: Objective
There is an increasing number of cuffless blood pressure (BP) measurement (BPM) devices. Despite promising results when comparing single measurements, the ability of these devices to track changes in BP levels over 24 h related to an initial calibration BP (CalibBP) is unknown. Our aim was to analyse this ability in a cuffless device using pulse transit time.Methods
We prospectively enrolled 166 participants for simultaneously performed cuffless (Somnotouch-NIBP) and cuff-based (Spacelabs 90217A/IEM Mobil-O-graph) 24 h BPM. As CalibBP for the cuffless device, first cuff-based BP was used. As surrogate for changes in BP levels after the CalibBP, we used the difference between the CalibBP and mean 24 h, awake and asleep BP measured by the two devices. In addition, w