Effect of hyperbilirubinemia on the accuracy of continuous non-invasive hemoglobin measurements in liver transplantation recipients.


ABSTRACT: This study evaluated the effect of hyperbilirubinemia on the accuracy of continuous non-invasive hemoglobin (SpHb) measurements in liver transplantation recipients. Overall, 1465 SpHb and laboratory hemoglobin (Hb) measurement pairs (n = 296 patients) were analyzed. Patients were grouped into normal (< 1.2 mg/dL), mild-to-moderate (1.2-3.0 mg/dL), and severe (> 3.0 mg/dL) hyperbilirubinemia groups based on the preoperative serum total bilirubin levels. Bland-Altman analysis showed a bias of 0.20 (95% limit of agreement, LoA: - 2.59 to 3.00) g/dL, 0.98 (95% LoA: - 1.38 to 3.35) g/dL, and 1.23 (95% LoA: - 1.16 to 3.63) g/dL for the normal, mild-to-moderate, and severe groups, respectively. The four-quadrant plot showed reliable trending ability in all groups (concordance rate > 92%). The rat