Validity and reliability of measurement of peripheral oxygen saturation during the 6-Minute Walk Test in patients with systemic sclerosis.


ABSTRACT: Peripheral oxygen saturation (SpO2) using the fingers may have important limitations due to Raynaud's phenomenon and sclerodactyly in patients with systemic sclerosis (SSc). Sensors located at more central body positions may be more accurate as these as less prone to Raynaud attacks. To determine the validity and reliability of the SpO2 measured at the finger, forehead, and earlobe during the 6-Minute Walk Test (6MWT). Eighty two patients with SSc had an arterial line placed while performing the 6MWT. Peripheral oxygen saturation was simultaneously measured by finger, forehead, and earlobe sensors and compared to the arterial oxygen saturation (SaO2) measured before and after the 6MWT. 40 patients repeated the 6MWT one week later to determine re-test reliab