Off-the-Shelf Tibial Cone Sizes May Not Accommodate All Patients' Bone Morphology and May Lead to Cortical Breaches in Revision Total Knee Arthroplasty: A 3D Modeling Study.


ABSTRACT:

Background

In revision total knee arthroplasty, tibial cones have demonstrated improved longevity and reduced incidence of aseptic loosening. Several currently available "off-the-shelf" (OTS) cone systems may not have sizes to accommodate all patient bone morphologies.

Methods

Computed tomographies from one hundred primary total knee arthroplasty patients and dimensions of 4 OTS cones were obtained. Press-fit stems were positioned in 3D tibia models to fit the diaphyseal trajectory. Cones were positioned around the stem at 1, 6, and 13 mm resections measured from the trough of the medial tibial plateau, simulating proximal tibial cuts and bone loss. Tibias were examined for cortical breaching following modeled cone preparation.

Results

Increased rate of breaching was