Ontology highlight
ABSTRACT: Significance
Bedside cerebral blood flow (CBF) monitoring has the potential to inform and improve care for acute neurologic diseases, but technical challenges limit the use of existing techniques in clinical practice.Aim
Here, we validate the Openwater optical system, a novel wearable headset that uses laser speckle contrast to monitor microvascular hemodynamics.Approach
We monitored 25 healthy adults with the Openwater system and concurrent transcranial Doppler (TCD) while performing a breath-hold maneuver to increase CBF. Relative blood flow (rBF) was derived from changes in speckle contrast, and relative blood volume (rBV) was derived from changes in speckle average intensity.Results
A strong correlation was observed between beat-to-beat optical rBF and T
SUBMITTER: Favilla CG
PROVIDER: S-EPMC10923543 | biostudies-literature | 2024 Jan
REPOSITORIES: biostudies-literature