ABSTRACT: Background
It is unknown whether bioprostheses used for transcatheter aortic valve implantation will have similar long-term durability as those used for surgical aortic valve replacement. Repetitive mechanical stress applied to the valve leaflets, particularly during diastole, is the main determinant of structural valve deterioration. Leaflet mechanical stress cannot be measured in vivo. The objective of this in vitro/in silico study was thus to compare the magnitude and regional distribution of leaflet mechanical stress in old vs new generations of self-expanding (SE) vs balloon expandable (BE) transcatheter heart valves (THVs).Methods
A double activation simulator was used for in vitro testing of two generations of SE THV (Medtronic CoreValve 26 mm and EVOLUT PRO 26 mm) a