Inter-scanner Aβ-PET harmonization using barrel phantom spatial resolution matching.


ABSTRACT:

Introduction

The standardized uptake value ratio (SUVR) is used to measure amyloid beta-positron emission tomography (Aβ-PET) uptake in the brainDifferences in PET scanner technologies and image reconstruction techniques can lead to variability in PET images across scanners. This poses a challenge for Aβ-PET studies conducted in multiple centers. The aim of harmonization is to achieve consistent Aβ-PET measurements across different scanners. In this study, we propose an Aβ-PET harmonization method of matching spatial resolution, as measured via a barrel phantom, across PET scanners. Our approach was validated using paired subject data, for which patients were imaged on multiple scanners.

Methods

In this study, three different PET scanners were evaluated: the Siemens Biograph

SUBMITTER: Ruwanpathirana GP 

PROVIDER: S-EPMC10927914 | biostudies-literature | 2024 Jan-Mar