Inter- and intra-rater reliability of knee flexion angle measurements on X-ray and MRI.


ABSTRACT:

Background

Range of motion (ROM) is an important aspect of orthopaedic patient assessment. It can be measured at the knee joint by determining the knee flexion angle (KFA) a patient can achieve at extremes of flexion and extension. As with any measurement, the accuracy and reliability of the method used determine its validity. The consistency of magnetic resonance imaging (MRI) scans as compared to the current gold standard of X-ray remains unknown in terms of KFA evaluation. The aim of this study was to assess and compare the reliability of measuring KFA between X-ray and MRI scans.

Methods

This study included 80 patients (94 knees) who had attended a specialist knee clinic due to varying knee pathologies and undergone both X-ray and MRI scans. Lateral and T1-weighted sagit

SUBMITTER: Summers HKC 

PROVIDER: S-EPMC10929294 | biostudies-literature | 2022