An unconstrained tripolar hip replacement matches a large capacity two-piece metal/polyethylene acetabular component with a bipolar prosthesis. This combination of accepted technology is different than the relatively new dual mobility prosthesis. The goal is to protect against dislocation and allow close to a normal range of motion (ROM). So far there has not been enough information about wear and performance of tripolar hip replacement to support its wide use.Methods
Twenty-four tripolar prostheses were retrieved from 23 patients after a mean of 14 years (range, 5-21 years). All implants had been placed in high-demand patients who participated in adventure sports, had occupations where a dislocation would be dangerous, or in patients undergoing revision. The tri