High early failure rate for a new unicondylar knee system.


ABSTRACT:

Background

This single-center retrospective study evaluated early failure rates for an unicompartmental knee arthroplasty (UKA) system with an anti-allergic surface.

Methods

We studied 87 consecutive joints received an UKA at a single center between 2017 and 2020. All patients received a fully cemented anti-allergic Univation-Aesculap partial knee replacement implant with a corundum blasting surface. All joints had precise indication of unicompartmental arthroplasty according to the current criteria of this procedure. The current series was restricted to patients undergoing medial cemented UKA. Medial compartment osteoarthritis was the main indication.

Results

We found early failure (aseptic loosening) was documented in 20 of the 87 joints (23%). The time to failure r