ABSTRACT: Purpose
Regular receiving coil quality assurance (QA) is required to ensure image quality of an MRIdian Linac system. The manufacturer provides a spherical phantom and positioning tube for single-slice signal-to-noise ratio (SNR) and uniformity assessments. We aimed to improve imaging setup and coverage and eliminate inter-scan variability by employing multi-slice imaging of a stable phantom. Additionally, we strived to expedite analysis by developing objective, automated analysis software.Methods
A 5300 mL cylindrical plastic bottle placed in plastic bins was scanned at isocenter using a spin-echo sequence with NEMA-recommended parameters and 18 axial slices, avoiding phantom repositioning. Acquisition was repeated with and without prescan normalization filtering and by sa