Validation of ToucHb, a non-invasive haemoglobin estimation: Effective for normal ranges, needs improvement for anaemia detection.


ABSTRACT: Non-invasive methods for haemoglobin estimation hold enormous potential for early detection and treatment of anaemia, especially in limited resource settings. We sought to validate the diagnostic accuracy of ToucHb, a non-invasive haemoglobin estimation device available in the Indian market. We prospectively evaluated the diagnostic performance of the ToucHb device using the Automated complete blood count (CBC) method as the gold standard. Persons referred for haemoglobin estimation to the central laboratory of the government medical college hospital in Mysore, southern India were included in the study. Out Of 140 people approached, 127 gave consent; 65% (n = 82) were female with median age of 37 (IQR 28-45). ToucHB reported median haemoglobin value of 14 g/dL compared to 13.3 g/dL for CBC