Fluorescence Spectra of Prototropic Forms of Fluorescein and Some Derivatives and Their Potential Use for Calibration-Free pH Sensing.
Ontology highlight
ABSTRACT: Fluorescence pH sensing has proven to be efficient but with the drawback that molecules photobleach, requiring frequent calibrations. Double-emission peak molecules allow ratiometric measurements and theoretically avoid calibration. However, they are often expensive and fragile and usually have very low quantum yields. Single emission peaks such as fluorescein and derivatives are inexpensive and have very high quantum yields. Because they are single emission peaks, the pH is assumed to be derived from the ratio of emitted intensities at measured pH and at high pH values, i.e., they require frequent calibration. However, the shape of their single emitted peak evolves slightly with pH. In this paper, we first demonstrate a simple method to calculate the emission spectrum shape of each protot
SUBMITTER: Gauthier-Manuel B
PROVIDER: S-EPMC10934543 | biostudies-literature | 2024 Mar
REPOSITORIES: biostudies-literature