ABSTRACT: Background
There are limited data regarding treatment for failed balloon-expandable transcatheter heart valves (THVs) in redo-transcatheter aortic valve implantation (TAVI).Aims
We aimed to assess THV performance, neoskirt height and expansion when performing redo-TAVI with the ACURATE platform inside a SAPIEN 3 (S3) compared to redo-TAVI with an S3 in an S3.Methods
Redo-TAVI was performed on the bench using each available size of the S3, the ACURATE neo2 (ACn2) and the next-generation ACURATE Prime XL (AC XL) implanted at 2 different depths within 20 mm/23 mm/26 mm/29 mm S3s serving as the "failed" index THV. Hydrodynamic testing was performed to assess THV function. Multimodality assessment was performed using photography, X-ray, microcomputed tomography (micro-CT
SUBMITTER: Meier D
PROVIDER: S-EPMC10941669 | biostudies-literature | 2024 Mar