ABSTRACT: Objectives
To evaluate torque maintenance and structural damage in implant components of different diameters subjected to a fatigue challenge.Methods
Thirty 10-mm-long, morse taper connection, titanium dental implants and their corresponding one-piece abutments were divided into three groups (n = 10) according to implant diameter: 4.3 mm (I4.3), 3.5 mm (I3.5), and 2.9 mm (I2.9). The implants were placed into a load-bearing fixture simulating bone tissue (modified G10), and the abutments were screwed into the implants to a final torque of 20 Ncm for the I4.3 and I3.5 and 15 Ncm for I2.9. The torque was secured by a digital torque meter. Cone-beam computed tomography (CBCT) scans were acquired and post-processed (e-Vol DX software) for all implant/abutment sets before and aft