Unknown

Dataset Information

Heart rate sensitivity of virtual non-contrast calcium scores derived from photon counting detector CT data: a phantom study.


ABSTRACT:

Purpose

To assess the reliability of virtual non-contrast (VNC) derived coronary artery calcium quantities in relation to heart rate and the VNC algorithm used compared to reference true non-contrast (TNC), considering several clinically established acquisition modes.

Material and methods

An ad hoc built coronary phantom containing four calcified lesions and an iodinated lumen was scanned using three cardiac acquisition modes three times within an anthropomorphic cardiac motion phantom simulating different heart rates (0, 60, 80, 100 bpm) and reconstructed with a conventional (VNCconv) and a calcium-sensitive (VNCpc) VNC algorithm. TNC reference was scanned at 0 bpm with non-iodinated lumen. Calcium scores were assessed in terms of number of lesions det

SUBMITTER: Risch F 

PROVIDER: S-EPMC10943147 | biostudies-literature | 2024 Mar

REPOSITORIES: biostudies-literature