Investigation of the mechanisms for wireless nerve stimulation without active electrodes.
Ontology highlight
ABSTRACT: Electric-field stimulation of neuronal activity can be used to improve the speed of regeneration for severed and damaged nerves. Most techniques, however, require invasive electronic circuitry which can be uncomfortable for the patient and can damage surrounding tissue. A recently suggested technique uses a graft-antenna-a metal ring wrapped around the damaged nerve-powered by an external magnetic stimulation device. This technique requires no electrodes and internal circuitry with leads across the skin boundary or internal power, since all power is provided wirelessly. This paper examines the microscopic basic mechanisms that allow the magnetic stimulation device to cause neural activation via the graft-antenna. A computational model of the system was created and used to find that under m
SUBMITTER: Smith LA
PROVIDER: S-EPMC10947236 | biostudies-literature | 2023 Oct-Dec
REPOSITORIES: biostudies-literature