Mitigating Matrix Effects in LC-ESI-MS-MS Analysis of a Urinary Biomarker of Xylenes Exposure.
Ontology highlight
ABSTRACT: Liquid chromatography-electrospray ionization-tandem mass spectrometry (LC-ESI-MS-MS) with stable isotope-labeled internal standards (SIL-ISs) is the gold standard for quantitative analysis of drugs and metabolites in complex biological samples. Significant isotopic effects associated with deuterium labeling often cause the deuterated IS to elute at a different retention time from the target analyte, diminishing its capability to compensate for matrix effects. In this study, we systematically compared the analytical performance of deuterated (2H) SIL-IS to non-deuterated (13C and 15N) SIL-ISs for quantifying urinary 2-methylhippuric acid (2MHA) and 4-methylhippuric acid (4MHA), biomarkers of xylenes exposure, with an LC-ESI-MS-MS assay. Analytical method comparison between ISs demonstrated