Triple-tuned birdcage and single-tuned dipole array for quadri-nuclear head MRI at 7 T.


ABSTRACT:

Purpose

The purpose of this work was to design and build a coil for quadri-nuclear MRI of the human brain at 7 T.

Methods

We built a transmit/receive triple-tuned (45.6 MHz for  2$$ {}^2 $$ H, 78.6 MHz for  23$$ {}^{23} $$ Na, and 120.3 MHz for  31$$ {}^{31} $$ P) quadrature four-rod birdcage that was geometrically interleaved with a transmit/receive four-channel dipole array (297.2 MHz for  1$$ {}^1 $$ H). The birdcage rods contained passive, two-pole resonant circuits that emulated capacitors required for single-tuning at three frequencies. The birdcage assembly also included triple-tuned matching networks, baluns, and transmit/receive switches. We assessed the performance of the coil with quality factor (Q) and signal-to-noise ratio (SNR) measurements, and performed in vivo multinuclear MRI and MR spectroscopic imaging (MRSI).

Results

Q measurements showed that the triple-tuned birdcage efficiency was within 33% of that of single-tuned baseline birdcages at all three frequencies. The quadri-tuned coil SNR was 78%, 59%, 44%, and 48% lower than that of single or dual-tuned reference coils for  1$$ {}^1 $$ H,  2$$ {}^2 $$ H,  23$$ {}^{23} $$ Na, and  31$$ {}^{31} $$ P, respectively. Quadri-nuclear MRI and MRSI was demonstrated in brain in vivo in about 30 min.

Conclusion

While the SNR of the quadruple tuned coil was significantly lower than dual- and single-tuned reference coils, it represents a step toward truly simultaneous quadri-nuclear measurements.

SUBMITTER: Paska J 

PROVIDER: S-EPMC10950522 | biostudies-literature | 2024 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Triple-tuned birdcage and single-tuned dipole array for quadri-nuclear head MRI at 7 T.

Paška Jan J   Wang Bili B   Chen Anna M AM   Madelin Guillaume G   Brown Ryan R  

Magnetic resonance in medicine 20231220 5


<h4>Purpose</h4>The purpose of this work was to design and build a coil for quadri-nuclear MRI of the human brain at 7 T.<h4>Methods</h4>We built a transmit/receive triple-tuned (45.6 MHz for  2$$ {}^2 $$ H, 78.6 MHz for  23$$ {}^{23} $$ Na, and 120.3 MHz for  31$$ {}^{31} $$ P) quadrature four-rod birdcage that was geometrically interleaved with a transmit/receive four-channel dipole array (297.2 MHz for  1$$ {}^1 $$ H). The birdcage rods contained passive, two-pole resonant circuits that emula  ...[more]

Similar Datasets

| S-EPMC10835210 | biostudies-literature
| S-EPMC5910212 | biostudies-literature
| S-EPMC12835457 | biostudies-literature
| S-EPMC4988950 | biostudies-literature
| S-EPMC9812758 | biostudies-literature
| S-EPMC7978235 | biostudies-literature
| S-EPMC2327287 | biostudies-literature
| S-EPMC9934596 | biostudies-literature
| S-EPMC10733907 | biostudies-literature
| S-EPMC11625660 | biostudies-literature