ABSTRACT: Introduction
Screening for chronic kidney disease relies on accurate and precise creatinine measurements. Traditionally, creatinine is measured in serum or plasma using high-throughput chemistry analyzers. However, dried blood spots (DBS) can also be utilized to improve testing access.Methods
Samples were obtained from a 6 mm DBS punch, which was reconstituted in water before undergoing an acetonitrile crash. The resulting supernatant was diluted using an 80:20 acetonitrile: water before injection. Creatinine was identified using an isocratic gradient, and detected using an API 4000 triple quadrupole mass analyzer. Quantification relied on matrix-matched calibrators, with values harmonized to the Roche Cobas enzymatic assay. Validation studies assessing method performance i