ABSTRACT: Purpose
Conventional breast MRI is performed in the prone position with a dedicated coil. This allows high-resolution images without breast motion, but the patient position is inconsistent with that of other breast imaging modalities or interventions. Supine breast MRI may be an interesting alternative, but respiratory motion becomes an issue. Motion correction methods have typically been performed offline, for instance, the corrected images were not directly accessible from the scanner console. In this work, we seek to show the feasibility of a fast, online, motion-corrected reconstruction integrated into the clinical workflow.Methods
Fully sampled T2 -weighted (T2 w) and accelerated T1 -weighted (T1 w) breast supine MR images we