ABSTRACT: Background
Measurements of CD4 T cells and hemoglobin (Hb) are conventionally used to determine the immunological state and disease progression for HIV-infected patients. We obtained a small lightweight point-of-care device, the BD FACSPrestoTM in order to demonstrate its ability to deliver CD4 and Hb analysis in comparison with two larger clinical machines the BDFACSCantoTM analyzer and Sysmex XN 1000 haematology analyzer. The advantages of using the POC device include access to HIV patient data in remote and in resource limited settings.Method
The analytical performance of the BD FACSPrestoTM, compared with the FACSCantoTM II flow cytometer and the Sysmex XN 1000 haematology analyzer was evaluated by testing 241 routine clinical specimens collected in EDTA tubes from pati