Mid-term results of a cementless hip stem in femoral revision: how much diaphyseal press-fit do we need?


ABSTRACT:

Introduction

The aim of this retrospective study was to analyze the clinical and functional outcome of a modular tapered revision hip stem after mid-term follow-up with a special focus on the length of the distal bicortical fixation of the cementless hip stem.

Materials and methods

Follow-up examination was carried out for all patients with implantation of the Prevision hip stem between 2014 and 2019 to collect demographic, functional, and radiographic data.

Results

44 patients with stem in situ were examined, and 61 patients could be included in the Kaplan-Meier survival analysis. Oxford's hip score was 37.3 at the mean follow-up of 4.0 years. Two hip stem revisions were performed due to periprosthetic infection, which resulted in a hip stem survival rate of 96.7% (C

SUBMITTER: Herold D 

PROVIDER: S-EPMC10965576 | biostudies-literature | 2024 Apr