Fast-Response Variable-Stiffness Magnetic Catheters for Minimally Invasive Surgery.
Ontology highlight
ABSTRACT: In minimally invasive surgery, such as cardiac ablation, magnetically steered catheters made of variable-stiffness materials can enable higher dexterity and higher force application to human tissue. However, the long transition time between soft and rigid states leads to a significant increase in procedure duration. Here, a fast-response, multisegmented catheter is described for minimally invasive surgery made of variable-stiffness thread (FRVST) that encapsulates a helical cooling channel. The rapid stiffness change in the FRVST, composed of a nontoxic shape memory polymer, is achieved by an active cooling system that pumps water through the helical channel. The FRVST displays a 66 times stiffness change and a 26 times transition enhancement compare with the noncooled version. The cathete
SUBMITTER: Piskarev Y
PROVIDER: S-EPMC10966541 | biostudies-literature | 2024 Mar
REPOSITORIES: biostudies-literature