Unknown

Dataset Information

Reproducibility and location-stability of radiomic features derived from cone-beam computed tomography: a phantom study.


ABSTRACT:

Objectives

This study aims to determine the reproducibility and location-stability of cone-beam computed tomography (CBCT) radiomic features.

Methods

Centrifugal tubes with six concentrations of K2HPO4 solutions (50, 100, 200, 400, 600, and 800 mg ml-1) were imaged within a customized phantom. For each concentration, images were captured twice as test and retest sets. Totally, 69 radiomic features were extracted by LIFEx. The reproducibility was assessed between the test and retest sets. We used the concordance correlation coefficient (CCC) to screen qualified features and then compared the differences in the numbers of them under 24 series (four locations groups * six concentrations). The location-stability was assessed using the Kruskal-Wal

SUBMITTER: He X 

PROVIDER: S-EPMC10968769 | biostudies-literature | 2023 Nov

REPOSITORIES: biostudies-literature