ABSTRACT: Background
The performance of high-frequency oscillatory ventilators (HFOV) differs by the waveform generation mode and circuit characteristics. Few studies have described the performance of piston-type HFOV. The present study aimed to compare the amplitude required to reach the target high-frequency tidal volume ([Formula: see text]); determine the relationship between the settings and actual pressure in amplitude or mean airway pressure ([Formula: see text]); and describe the interaction among compliance, frequency, and endotracheal tube (ETT) inner diameter in 4 HFOV models, including Humming X, Vue (a piston type ventilator commonly used in Japan), VN500 (a diaphragm type), and SLE5000 (a reverse jet type).Methods
The oscillatory ventilators were evaluated by using a 50