ABSTRACT: Background
Lesion size is reported to become larger as contact force (CF) increases. However, this has not been systematically evaluated in temperature-guided very high-power short-duration (vHPSD) ablation, which was therefore the purpose of this study.Methods
Radiofrequency applications (90 W/4 s, temperature-control mode) were performed in excised porcine myocardium with four different CFs of 5, 15, 25, and 35 g using QDOT-MICRO™ catheter. Ten lesions for each combination of settings were created, and lesion metrics and steam-pops were compared.Results
A total of 320 lesions were analyzed. Lesion depth, surface area, and volume were smallest for CF of 5 g than for 15, 25, and 35 g (depth: 2.7 mm vs. 2.9 mm, 3.0 mm, 3.15 mm, p < .01; surface area: 38.4 mm