Determination of carbon in microplastics and single cells by total consumption microdroplet ICP-TOFMS.


ABSTRACT: Recently, spICP-MS analyses of microplastics have demonstrated that the detection capabilities of ICP-MS are sufficient to determine the size and composition of such materials. However, solution nebulization or microdroplet generation limits the sizes of droplets, microparticles, or cells that can be efficiently introduced into commonly used horizontal ICP-MS configurations. Therefore, we introduced the development of a downward-pointing ICP coupled to a time-of-flight mass spectrometer (ICP-TOFMS), which enables quantitative transport of large microdroplets (diameters up to 90 µm) into the ICP. Here, we report the capabilities of downward ICP-TOFMS for the quantitative analysis of single cells and microplastic particles. For calibration of element mass amount per particle or cell, microdr