Unknown

Dataset Information

Real-world validation of smartphone-based photoplethysmography for rate and rhythm monitoring in atrial fibrillation.


ABSTRACT:

Aims

Photoplethysmography- (PPG) based smartphone applications facilitate heart rate and rhythm monitoring in patients with paroxysmal and persistent atrial fibrillation (AF). Despite an endorsement from the European Heart Rhythm Association, validation studies in this setting are lacking. Therefore, we evaluated the accuracy of PPG-derived heart rate and rhythm classification in subjects with an established diagnosis of AF in unsupervised real-world conditions.

Methods and results

Fifty consecutive patients were enrolled, 4 weeks before undergoing AF ablation. Patients used a handheld single-lead electrocardiography (ECG) device and a fingertip PPG smartphone application to record 3907 heart rhythm measurements twice daily during 8 weeks. The ECG was performed immediately b

SUBMITTER: Gruwez H 

PROVIDER: S-EPMC11023210 | biostudies-literature | 2024 Mar

REPOSITORIES: biostudies-literature