Validation of a novel smartphone-based photoplethysmographic method for ambulatory heart rhythm diagnostics: the SMARTBEATS study.


ABSTRACT:

Aims

In the current guidelines, smartphone photoplethysmography (PPG) is not recommended for diagnosis of atrial fibrillation (AF), without a confirmatory electrocardiogram (ECG) recording. Previous validation studies have been performed under supervision in healthcare settings, with limited generalizability of the results. We aim to investigate the diagnostic performance of a smartphone-PPG method in a real-world setting, with ambulatory unsupervised smartphone-PPG recordings, compared with simultaneous ECG recordings and including patients with atrial flutter (AFL).

Methods and results

Unselected patients undergoing direct current cardioversion for treatment of AF or AFL were asked to perform 1-min heart rhythm recordings post-treatment, at least twice daily for 30 days at