Ontology highlight
ABSTRACT: Purpose
To develop an iterative concomitant field and motion corrected (iCoMoCo) reconstruction for isotropic high-resolution UTE pulmonary imaging at 0.55 T.Methods
A free-breathing golden-angle stack-of-spirals UTE sequence was used to acquire data for 8 min with prototype and commercial 0.55 T MRI scanners. The data was binned into 12 respiratory phases based on superior-inferior navigator readouts. The previously published iterative motion corrected (iMoCo) reconstruction was extended to include concomitant field correction directly in the cost function. The reconstruction was implemented within the Gadgetron framework for inline reconstruction. Data were retrospectively reconstructed to simulate scan times of 2, 4, 6, and 8 min. Image quality was assessed using apparen
SUBMITTER: Javed A
PROVIDER: S-EPMC11055666 | biostudies-literature | 2024 Jul
REPOSITORIES: biostudies-literature