Ontology highlight
ABSTRACT: Purpose
We developed a novel transpalpebral self-tonometer called the TapEye tonometer (TET) based on palpation of the upper eyelid. Our goal was to evaluate a method for improving the accuracy of measuring intraocular pressure (IOP) through the eyelid.Methods
Participants underwent standardized training by technicians and were required to be able to use the TET for study inclusion. Subsequently, a noncontact tonometer and Goldmann applanation tonometer (GAT) were used. All participants were instructed to measure their IOPs using the three tonometers at baseline (visit 1) and at 1 month (visit 2). At visit 2, the corrected IOP value measured by the TET (c-TET) was calculated using the difference between the TET and GAT measurements obtained at visit 1.Results
No sig
SUBMITTER: Kobashi H
PROVIDER: S-EPMC11095731 | biostudies-literature | 2024
REPOSITORIES: biostudies-literature