Ontology highlight
ABSTRACT: Background
Compared with thin-strut durable-polymer drug-eluting stents (DP-DES), ultrathin-strut biodegradable-polymer sirolimus-eluting stents (BP-SES) improve stent-related clinical outcomes in patients undergoing percutaneous coronary intervention (PCI). Reduced stent strut thickness is hypothesised to underlie these benefits, but this conjecture remains unproven.Aims
We aimed to assess the impact of strut thickness on stent healing and clinical outcomes between ultrathin-strut and thin-strut BP-SES.Methods
First, we performed a preclinical study of 8 rabbits implanted with non-overlapping thin-strut (diameter/thickness 3.5 mm/80 μm) and ultrathin-strut (diameter/thickness 3.0 mm/60 μm) BP-SES in the infrarenal aorta. On day 7, the rabbits underwent intravascula
SUBMITTER: Ikegami R
PROVIDER: S-EPMC11100507 | biostudies-literature | 2024 May
REPOSITORIES: biostudies-literature