The force at the implant cannot be assessed by the mallet force-Unless supported by a model.
Ontology highlight
ABSTRACT: The implantation of uncemented prostheses requires the application of sufficient forces to achieve a press-fit of the implant in the bone. Excessive forces have to be omitted to limit bone damage. Force measurements along the force transmission path between mallet and implant are frequently used to investigate this trade-off. Placing a load cell at a position of interest (PoI), which might be the implant bone interface or the head taper junction, is technically challenging or even impossible so that nearby positions are chosen. Thus, a certain inertia and stiffness remain between the PoI and the sensor, and consequently the measured dynamic forces differ from those at the PoI. This experimental and numerical study aimed to investigate the amount of force reduction along the transmission pa
SUBMITTER: Schlieker PJ
PROVIDER: S-EPMC11101067 | biostudies-literature | 2024