Technique for Recanalization of Sheath Introducer Occlusion due to Captured Thrombus during Mechanical Thrombectomy for Acute Ischemic Stroke: A Technical Note.


ABSTRACT:

Objective

Mechanical thrombectomy (MT) is the gold standard treatment for acute ischemic stroke. During these interventions, a thrombus frequently obstructs a guiding catheter. The obstructed guiding catheter should be withdrawn before distal embolism occurs; however, albeit infrequently, the thrombus occludes even a sheath introducer (SI). While conventionally new SI placement is required for continuation of treatment, we propose a viable alternative for recanalization of the occluded SI, termed vacuum-assisted delivery of thrombus (VADT), with a clinical report of our cases. The usefulness of this technique was also evaluated in simulation experiments.

Case presentations

The VADT procedure is as follows: 1) insert a peel-away sheath, originally attached to a balloon-guidin