Validation of Signal Intensity Gradient from TOF-MRA for Wall Shear Stress by Phase-Contrast MR.
Ontology highlight
ABSTRACT: To validate the correlation between the signal intensity gradient (SIG) from time-of-flight magnetic resonance angiography (TOF-MRA) and wall shear stress (WSS) determined by phase contrast magnetic resonance (PC-MR), we conducted both experimental and human studies. In the experimental study, we measured WSS in four tubes of different sizes with variable flow rates using PC-MR and TOF-MRA. The flow rates of water in the experimental study ranged from 0.06 to 12.75 mL/s, resulting in PC-WSS values between 0.1 and 1.6 dyne/cm2. The correlation between PC-WSS and SIG was statistically significant, showing a coefficient of 0.86 (P < 0.001, R2 = 0.75). The line fit provided the conversion equation as Y = 1.6287X - 1.1563 (Y = PC-WSS, X = SIG). For the human study, 28 subj
SUBMITTER: Lee CH
PROVIDER: S-EPMC11169296 | biostudies-literature | 2024 Jun