Fast and accurate newborn heart rate monitoring at birth: A systematic review.


ABSTRACT:

Aim

To examine speed and accuracy of newborn heart rate measurement by various assessment methods employed at birth.

Methods

A search of Medline, SCOPUS, CINAHL and Cochrane was conducted between January 1, 1946, to until August 16, 2023. (CRD 42021283364) Study selection was based on predetermined criteria. Reviewers independently extracted data, appraised risk of bias and assessed certainty of evidence.

Results

Pulse oximetry is slower and less precise than ECG for heart rate assessment. Both auscultation and palpation are imprecise for heart rate assessment. Other devices such as digital stethoscope, Doppler ultrasound, an ECG device using dry electrodes incorporated in a belt, photoplethysmography and electromyography are studied in small numbers of newborns and d