Unknown

Dataset Information

Accuracy and variability of cardiologist interpretation of single lead electrocardiograms for atrial fibrillation: The VITAL-AF trial.


ABSTRACT:

Background

Screening for atrial fibrillation (AF) using consumer-based devices capable of producing a single lead electrocardiogram (1L ECG) is increasing. There are limited data on the accuracy of physician interpretation of these tracings. The goal of this study is to assess the sensitivity, specificity, confidence, and variability of cardiologist interpretation of point-of-care 1L ECGs.

Methods

Fifteen cardiologists reviewed point-of-care handheld 1L ECGs collected from patients aged 65 years or older enrolled in the VITAL-AF clinical trial [NCT035115057] who underwent cardiac rhythm assessments with a 1L ECG using an AliveCor KardiaMobile device. Random sampling of 1L ECGs for cardiologist review was stratified by the AliveCor algorithm interpretation. A 12L ECG performe

SUBMITTER: Pipilas DC 

PROVIDER: S-EPMC11194686 | biostudies-literature | 2023 Nov

REPOSITORIES: biostudies-literature