Development of a Tetherless Bioimpedance Device That Uses Morphologic Changes to Predict Blood Flow Restrictions Mimicking Peripheral Artery Disease Progression.
Ontology highlight
ABSTRACT: A tetherless multi-targeted bioimpedance device was designed, modeled, built, and tested for measuring arterial pulse and, using morphological analysis, its potential for monitoring blood flow restrictions that mimic Peripheral Artery Disease (PAD) was assessed across multiple peripheral arteries. Specifically, we first developed a small form factor, tetherless, bioimpedance device, based on high-frequency structure simulator (HFSS) simulations. After designing and building the device we then tested it in vivo on human subjects on multiple arteries and found that we did not need to modify the gain on the device compared to the bench top system. Further, it was found that changes in the morphology of the bioimpedance signal over time, depicted through the ratio of the first and second harmo