ABSTRACT: Objectives
Despite clinical guidelines do not recommend the use of point-of-care testing (POCT) glucometers for diagnostic purposes yet, the analytical performance is continuously improving. Thus, we evaluate the technical accuracy and clinical concordance of POCT glucometers during an oral glucose tolerance test (OGTT) in children for prediabetes and diabetes diagnosis in a comparison study.Methods
Pediatric patients with an OGTT indication who attended the Diabetes Unit between December 2020 and September 2021 were recruited for this prospective observational study. During the functional test, glycaemia was immediately measured in venous blood using two glucometers (unconnected and connected) and sent to the central laboratory.Results
The study included 98 patient