ABSTRACT: Purpose
To assess the accuracy of a prototype novel instrument for intra ocular pressure (IOP) measurements not involving corneal pressure application.Design
Prospective case control study.Methods
An institutional study including 16 healthy volunteers without ocular pathology. IOP in both eyes of the participants was measured four times in different body positions with the novel prototype and reference instrument (Goldmann applanation tonometer (GAT) or iCare (iCare Finland OY, Vantaa, Finland)). IOP results were compared between the prototype and the reference instruments in 116 pairs of measurement.Results
Overall no statistically significant difference was found between the presented prototype and the reference instrument. Stratifying measurements by inst