ABSTRACT: Purpose
To compare the accuracy of the Barrett II universal (BU II) formula, Hoffer-Q, and SRKT formulae following lensectomy and IOL implantation in a large pediatric cohort.Methods
Retrospective study of children who underwent lensectomy and IOL implantation between 2015 and 2023 at Hadassah-Hebrew University Medical Center, Jerusalem, Israel.Results
One hundred and fifty-one eyes of 104 children aged 6.0 ± 3.9 years were included. The mean prediction error (PE) was - 0.08 ± 1.54 diopters (D) with BU II, 0.24 ± 1.46 D with Hoffer-Q, and 0.71 ± 1.92 D with SRKT (P = 0.10). In eyes with axial length (AL) < 22 mm, BU II and Hoffer-Q had a smaller PE than SRKT (P = 0.024). In eyes with AL ≥ 22 mm, BU II had a smaller PE than Hoffer-Q (P = 0.048). In children 24 months