Unknown

Dataset Information

Integration of pH Control into Chi.Bio Reactors and Demonstration with Small-Scale Enzymatic Poly(ethylene terephthalate) Hydrolysis.


ABSTRACT: Small-scale bioreactors that are affordable and accessible would be of major benefit to the research community. In previous work, an open-source, automated bioreactor system was designed to operate up to the 30 mL scale with online optical monitoring, stirring, and temperature control, and this system, dubbed Chi.Bio, is now commercially available at a cost that is typically 1-2 orders of magnitude less than commercial bioreactors. In this work, we further expand the capabilities of the Chi.Bio system by enabling continuous pH monitoring and control through hardware and software modifications. For hardware modifications, we sourced low-cost, commercial pH circuits and made straightforward modifications to the Chi.Bio head plate to enable continuous pH monitoring. For software integration,

SUBMITTER: Denton MCR 

PROVIDER: S-EPMC11223484 | biostudies-literature | 2024 Jul

REPOSITORIES: biostudies-literature