Multicenter validation study for automated left ventricular ejection fraction assessment using a handheld ultrasound with artificial intelligence.
Ontology highlight
ABSTRACT: We sought to validate the ability of a novel handheld ultrasound device with an artificial intelligence program (AI-POCUS) that automatically assesses left ventricular ejection fraction (LVEF). AI-POCUS was used to prospectively scan 200 patients in two Japanese hospitals. Automatic LVEF by AI-POCUS was compared to the standard biplane disk method using high-end ultrasound machines. After excluding 18 patients due to infeasible images for AI-POCUS, 182 patients (63 ± 15 years old, 21% female) were analyzed. The intraclass correlation coefficient (ICC) between the LVEF by AI-POCUS and the standard methods was good (0.81, p < 0.001) without clinically meaningful systematic bias (mean bias -1.5%, p = 0.008, limits of agreement ± 15.0%). Reduced LVEF < 50% was detected with a sensitivity of 85
SUBMITTER: Kagiyama N
PROVIDER: S-EPMC11224326 | biostudies-literature | 2024 Jul
REPOSITORIES: biostudies-literature
ACCESS DATA