ABSTRACT: Background
Therapeutic drug monitoring of infliximab (IFX) can improve treatment outcomes; however, the temporal gap between drug concentration monitoring and subsequent availability restricts its practical application. To address this issue, an automated monitoring method, AFIAS IFX, was developed to rapidly and accurately analyze IFX concentration in blood. The analytical and clinical performances of this method were assessed to establish its clinical utility.Methods
The analytical performance of AFIAS IFX was evaluated according to Clinical and Laboratory Standard Institute guidelines. For clinical validation, AFIAS IFX was compared with 3 established enzyme-linked immunosorbent assay kits (LISA TRACKER, RIDASCREEN, and ImmunoGuide) using 100 consecutive samples from 28