Green method for 17-hydroxyprogesterone extraction and determination using PDMS stir bar sorptive extraction coupled with HPLC: optimization by response surface methodology.


ABSTRACT: Quantifying small amounts of the 17-hydroxyprogesterone in various matrix is crucial for different purposes. In this study, a commercial polydimethylsiloxane stir bar was used to extract hormone from water and urine samples. Analysis was performed by high-performance liquid chromatography using a UV detector. The response surface methodology was used to optimize the desorption and extraction steps, with predicted optimal point relative errors of 1.25% and 6.40%, respectively. The optimized method was validated with a linear range of 1.21-1000.00 for aqueous and 2.43-2000.00 ng mL-1 for urine samples. The coefficient of determination was 0.9998 and 0.9967, and the detection limit of the proposed method was obtained to be 0.40 and 0.80 ng mL-1 for aqueous and urine samp