ABSTRACT: Background
Monitoring blood oxygenation is essential in immobilised rhinoceros, which are susceptible to opioid-induced hypoxaemia. This study assessed the reliability, clinical performance and trending ability of the Nonin PalmSAT 2500 A pulse oximeter's and the Masimo Radical-7 pulse co-oximeter's dual-wavelength technology, with their probes placed at two measurement sites, the inner surface of the third-eyelid and the scarified ear pinna of immobilised white rhinoceroses. Eight white rhinoceros were immobilised with etorphine-based drug combinations and given butorphanol after 12 min, and oxygen after 40 min, of recumbency. The Nonin and Masimo devices, with dual-wavelength probes attached to the third-eyelid and ear recorded arterial peripheral oxygen-haemoglobin saturation (S