Clip opening while locked after transcatheter edge-to-edge mitral valve repair with different onset times: a case series.


ABSTRACT:

Background

Transcatheter edge-to-edge mitral valve repair is now available in many countries and has achieved favourable therapeutic outcomes. However, there have been no reported cases of clip opening while locked (COWL) during the acute phase using the MitraClip G4 system (Abbott, Abbott Park, IL, USA).

Case summary

We present two cases of COWL occurring at different phases: one immediately after clip release and the other 2 days post-procedure. In both cases, the initial treatment involved the use of the XTW system. Subsequently, an additional XT system was deployed for the deterioration of mitral regurgitation caused by COWL, without any complications.

Discussion

The MitraClip G4 system offers four size variations, providing a larger grasping area and increased fl