Efficacy and safety of focal pulsed-field ablation for ventricular arrhythmias: two-centre experience.


ABSTRACT:

Aims

A pulsed electric field (PF) energy source is a novel potential option for catheter ablation of ventricular arrhythmias (VAs) as it can create deeper lesions, particularly in scarred tissue. However, very limited data exist on its efficacy and safety. This prospective observational study reports the initial experience with VA ablation using focal PF.

Methods and results

The study population consisted of 44 patients (16 women, aged 61 ± 14years) with either frequent ventricular premature complexes (VPCs, 48%) or scar-related ventricular tachycardia (VT, 52%). Ablation was performed using an irrigated 4 mm tip catheter and a commercially available PF generator. On average, 16 ± 15 PF applications (25 A) were delivered per patient. Acute success was achieved in 84% of pati

SUBMITTER: Peichl P 

PROVIDER: S-EPMC11264298 | biostudies-literature | 2024 Jul